Waspmote Nitrogen Dioxide Sensor

Quick Data

Conversion Method

Converting to Concentration in Air (parts per million)

The output of the waspmote to the database is a voltage reading, as can be seen from the circuit diagram below, this reading is affected by the input voltage, sensor resistance and load resistance.

To convert to ppm, the voltage reading is applied as follows

  1. Convert voltage reading to sensor resistance: Rsensor = Rload / ( Vcc - Vout) * Vout
  2. Normalise against resistance under air: Rsensor / Ro(Typical Resistance in Air)
  3. Compare normalised result against sensor response diagram:

Sensor Response Characteristics

Normalised Resistance (Rs/Ro) vs NO2 (ppm)
NO2 (ppm) = 10(log(Rs/Ro) - 2.702) / 1.865

Example

Measurement of 1.277v is observed (from our databases)